LaCoste
& Romberg LLC
The first name in gravity
since 1939
MVR FEEDBACK
SYSTEM MANUAL
CHAPTER 5
FEEDBACK OUTPUT MONITORING
INTERNAL 4.5 DIGIT SYSTEM MONITOR
You can monitor the Feedback Voltage output using the
INTERNAL 4.5-Digit SYSTEM MONITOR. Simply set the selector switch
to the position marked "MVR OUTPUT"
The internal System Monitor
DVM is installed inside the lid of the gravity meter in place of
the thermometer. It uses special low-power, high-intensity red
LEDs that function in both low and high temperature situations.
The DVM indicates the
voltage output of the Feedback card in millivolts. A filter
before the DVM provides about 60 seconds of averaging. The
natural 1 to 2 second oscillation of the MVR Feedback voltage is
thereby averaged to a steady value that can be recorded by the
operator.
EXTERNAL DVM MONITOR
A battery-powered DVM can be used (Fluke Model 85 or similar
unit) for measuring the Feedback voltage. The DVM is connected to
the Analog Output Connector ("Recorder Jack") on the
side of the gravity meter.
It is helpful to use the AVERAGE feature when measuring the
Feedback voltage using an external DVM. The voltage can also be
recorded by a data acquisition system if desired.
Table of
Contents
Chapter
6
Product Information
Copyright
© 1998 LaCoste and Romberg LLC